 |
PINPOINT¨ Nitrate Monitor
User's Guide
American Marine is proud to introduce the PINPOINT¨ Nitrate Monitor. The only truly affordable and accurate digital Nitrate measurement instrument for marine saltwater. The PINPOINT¨ Nitrate Monitor
has been designed as a precision spot-testing instrument. A 9-volt
battery is recommended. Replace battery when ÒLOBATÓ appears on the
display.
NOTE: The
Nitrate Monitor unit is NOT waterproof and must be operated on a dry
surface. Liquid contact on the printed circuit board will cause
corrosion and void warranty.
Set-up
Unwrap the plastic bubblewrap packaging from both the meter and
the probe. Carefully unscrew and remove the plastic storage bottle from
the end of the probe. You may leave the bottle top on the probe. DO NOT
discard the storage bottle, it will be used for calibration, measurement
and storage. Remove the battery cover on the back of the monitor and
install a 9-volt battery. Attach the BNC connector to the input on the
front of the monitor, and rotate 1/4 turn clockwise to lock the
connector.
New Probe
Condition a new probe by putting the tip into each calibration
fluid 10ppm, 100ppm and 1ppm for 20 minutes in each fluid. This is
usually only necessary for the first time use.
High Range Calibration (10 ppm to 100 ppm)
- Pour 10 ppm Nitrate Calibration Fluid
into the storage bottle to the fill line. Insert the end of the probe
into the fluid and screw the top onto the bottle. Check the flat probe
tip is not resting on the bottom of the bottle.
- Slide the Range Switch to the HIGH position
- Press and release the ON/OFF button to power on the monitor.
- Press and release the Calibrate button. ÒCALÓ will appear
on the display and the number 10 will flash slowly for about 40
seconds.
- When the number 100 starts to flash rapidly, the meter is
ready for the 100 ppm fluid. DonÕt rush; you have as much time as you
need. Unscrew the bottle and return the 10 ppm calibration fluid back
into its original bottle. Pour 100 ppm nitrate calibration fluid into
the storage bottle to the fill line. Insert the end of the probe into
the fluid and attach the bottle.
- Press and release the Calibrate button. The number 100
will flash slowly for about 40 seconds. When the flashing stops the
meter is fully calibrated for the HIGH Range. Pour the 100 ppm
calibration fluid back into the original container.
- Pour a water sample to the fill line on the storage
bottle and insert the probe to determine the exact Nitrate level. Allow
up to one minute for the reading to stabilize. When finished, rinse the
probe with R/O or deionized water and blot dry with a paper towel prior
to storage.
Low Range Calibration ( 1 ppm to 10 ppm)
- Pour 1 ppm Nitrate Calibration Fluid
into the storage bottle to the fill line. Insert the end of the probe
into the fluid and screw the top onto the bottle. Check the flat probe
tip is not resting on the bottom of the bottle.
- Slide the Range Switch to the LOW position
- Press and release the ON/OFF button to power on the monitor.
- Press and release the Calibrate button. ÒCALÓ will appear
on the display and the number 1 will flash slowly for about 40 seconds.
- When the number 10 starts to flash rapidly, the meter is
ready for the 10 ppm fluid. DonÕt rush; you have as much time as you
need. Unscrew the bottle and return the 1 ppm calibration fluid back
into its original bottle. Pour 10 ppm nitrate calibration fluid into the
storage bottle to the fill line. Insert the end of the probe into the
fluid and screw the top onto the bottle.
- Press and release the Calibrate button. The number 10
will flash slowly for about 40 seconds. When the flashing stops the
meter is fully calibrated for the LOW Range. Pour the 10 ppm calibration
fluid back into the original container.
- Pour a water sample to the fill line on the storage
bottle and insert the probe to determine the exact Nitrate level. Allow
up to one minute for the reading to stabilize. When finished, Rinse
probe with RO or deionized water and blot dry with a paper towel prior
to storage.
Important Points to Remember
- PINPOINT¨ Nitrate Monitor
has been designed as a high precision spot-testing monitor. Continuous
Nitrate Monitoring may not always be possible and may require more
frequent recalibration and probe replacement.
- PINPOINT¨ Nitrate Monitor
is NOT waterproof and must be operated on a dry surface. Liquid contact
with the printed circuit board may cause corrosion and void warranty.
- PINPOINT¨ Nitrate Monitor has 2 independent ranges:
0 Ð 10 ppm NO3-N (nitrate-nitrogen)
10Ð100 ppm NO3-N (nitrate-nitrogen)
- Accuracy for either scale after calibration is +/- 1 ppm NO3-N (nitrate-nitrogen)
- Resolution for either scale is 1 ppm NO3-N (nitrate-nitrogen)
- When the probe is not in use, the tip should be stored in
the protective bottle (dry). That is with NO fluid added to the bottle.
A properly stored probe may have a life up to 2Ð3years.
- For greatest accuracy all calibration and sample fluids should be at (the same) room temperature.
- The Nitrate Meter will always store the last calibration even when the power is turned off.
- PINPOINT¨ Nitrate Probe
should be kept very clean. Rinse off the probe tip with RO/DI water
prior to storage. Do not use any cleaning fluids or chemicals.
- Calibration fluids can be reused and should be replaced
occasionally. Calibration Fluid replacement will depend on individual
technique and be indicated by the inability to calibrate.
- For increased stability allow the probe tip to soak in calibration fluid for a minute or so before calibration.
- There are several possibilities for the meter to display ErÉP during the calibration process:
1-The probe BNC is not connected to the meter.
2-The probe membrane is very dry (common with a new probe). Soak
the probe tip in both 10 ppm or 100 ppm calibration fluid for 60 minutes
each fluid. Recalibrate.
3-The probe has reached the end of its useful life
- Lo on the display indicates the probe is measuring a nitrate value below the selected scale.
- High on the display indicates the probe is measuring a nitrate value above the selected scale.
Warranty
PINPOINT¨ Nitrate Monitor by
American Marine Inc. is warranted to be free of defects in material and
workmanship for a period of 2 years from date of sale.
PINPOINT¨ Nitrate Probe is warranted to be free of defects in material and workmanship for a period of 6 months from the date of sale.
Positive proof of purchase is required for warranty claim.
Removal or alteration of the serial number will void warranty.
American Marine Inc. will not be liable for any costs of removal,
installation, transportation charges, or any other charges, which may
result in connection with a warranty claim.
American Marine Inc. will not be liable for any damage or wear to
products or livestock caused by abnormal operating conditions, water
damage, abuse, misuse, unauthorized alternation or repair or if the
product was not installed in accordance with the printed operating
instructions.
Any defective product to must be sent freight prepaid with
appropriate documentation supporting the warranty claim. Replacement or
repair will be at the discretion of American Marine Inc. Typical
turnaround time within 48
hours. Overnight delivery available.
American Marine Inc.
54 Danbury Road, Suite 172
Ridgefield, CT 06877 U.S.A.
Fax/Phone: 914-763-5367
|
 |